-
-
Product NamePrince-100N Fingertip Pulse OximeterProduct Description
-5 seconds out of value. High-performance MCU, greatly increasing the computing speed
-lmported chips. Equipped with European imported chips for more accurate capturing of blood oxygen pulse signal
-Optical frequency digital conversion technology. Adopt the second generation LFC PLUS optical frequency digital conversion technology to greatly improve the measurement accuracy
-Audible alarm. Automatic alarm in case of exceeding the limit, the upper and lower limits of the alarm can be set by yourself
-Smart power saving. Automatic power-on monitoring after the finger is inserted into the machine aautomatically shut down after 5s of finger removal
Performance characteristics
1. Lightweight design, easy to carry, suitable for family use.
2. 1.06 inch color bright OLED display with humanized observation interface and short press the button. The screen can be rotated in four directions.
3. It can display blood oxygen saturation, pulse rate, blood perfusion index, pulse intensity histogram, pulse volume waveform.
4. Intelligent broadcast of blood oxygen saturation, pulse rate and abnormal pulse rate analysis result prompt
5. 12 sets of data storage with memory function
6. Equipped with spot measurement and real-time monitoring mode, automatic sound-light alarm triggerswhen exceeding the limit
7. Intelligent power saving mode
8. Good performance under low perfusion; strong anti-interference ability
9. Two AAA batteries, can be used continuously for 30 hours
Technical parameters
Blood oxygen saturation (SPO2)
Measuring range: 35%~100% Display range: 35%~99%
Measurement accuracy: 70%~100%: ±2%; 50%~69%: ±3%;
No requirement below 50% (where the percentage sign "%" is the pulse oximetry percentage)
Measurement accuracy: in the 70~100% measurement range, Arms≤3% (For the definition of Arms, please refer to the standard YY0784-2010)
Pulse rate (PR)
Measuring range/display range: 30bpm~240bpm;
Measurement error: ±2% or ±2bpm, whichever is greater
Perfusion Index (PI)
Measuring range: 0.2%~20%
Preset limit value
Prompt setting range when blood oxygen saturation exceeds the limit:
Lower limit prompt: adjustable within the range of 85%~99%, the default is 90%
Reminder range of pulse rate exceeding limit:
Upper limit: 100bpm~240bpm, the default is 120bpm;
Lower limit: 30bpm~60bpm, default 50bpm
Anti-interference
Anti-interference ability of natural light: When used in a natural light environment, the measured values of pulse oximetry and pulse rate meet the requirements of the above-mentioned standards.
Anti-power frequency interference capability: When the light environment is a power frequency light source, the pulse oximetry and pulse rate measurement values meet the requirements of the above standards.
Power supply
Working voltage: d.c.3V (2 AAA 1.5V alkaline batteries)
Current: ≤50mA
Specification
Dimensions: length 66 × width 36 × height 33 (mm)
Net weight: 60g (including battery)
Environmental requirements
Operating temperature: 5℃~40℃
Storage temperature: -20℃~55℃
Environmental humidity: Working humidity: 30%~80%, non-condensing
Storage humidity: 10%~100%
Atmospheric pressure: working range: 70kPa~106kPa
Storage range: 50kPa~106kPa
Contact us: export@healforce.com
Copyright © 2020 Heal Force.All Rights Reserved.